Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1022
User Manual 1178.8520.02 ─ 08
<PulsePeriod>
Defines the pulse period.
Default unit: s
<OfPulses>
Defines the number of pulses to measure.
Example:
CALC:MARK:FUNC:MSUM 50US,450US,576.9US,8
Evaluates data that contains 8 pulses during a measurement
time of 450
µs and a pulse period of 576.9
µs. The evaluation
starts with an offset of 50
µs.
[SENSe:]MPOWer:FTYPe
<FilterType>
This command selects the filter type for pulse power measurements.
Parameters:
<FilterType>
CFILter
NORMal
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FSV3-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FSV3-K54) option
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Return values:
<PulsePower>
List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Usage:
Query only
[SENSe:]MPOWer[:SEQuence]
<Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <NoPulses>
This command configures and initiates the pulse power measurement.
The R&S
FSV/A caches all measurement parameters that you can set with this com-
mand. If you use the command repeatedly, the R&S
FSV/A only changes those set-
tings that you have actually changed before initiating the measurement. Thus, mea-
surement times are kept as low as possible.
Configuring and performing measurements