Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1130
User Manual 1178.8520.02 ─ 08
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
PSEN
External power sensor
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
"Using the power sensor as an external trigger"
on page 367
See
See
See
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
numeric value
Range:
1 us to 15 s
*RST:
10 ms
Default unit: S
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 5
The sweep starts every 5 s.
Manual operation:
See
12.7.5.2
Configuring gated measurements
............................................................................................... 1131
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt
............................................................... 1131
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth
..............................................................1131
[SENSe:]SWEep:EGATe:CONTinuous[:STATe]
................................................................1132
..................................................................................1132
...................................................................................1132
..................................................................................1133
..................................................................................1133
......................................................................................1134
Setting basic measurement parameters