Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1104
User Manual 1178.8520.02 ─ 08
Example:
//Couple step size to span
FREQ:CENT:STEP:LINK SPAN
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
<Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor>
1 to 100 PCT
*RST:
10
Default unit: PCT
Example:
//Couple frequency step size to span and define a step size fac-
tor
FREQ:CENT:STEP:LINK SPAN
FREQ:CENT:STEP:LINK:FACT 20PCT
Manual operation:
See
[SENSe:]FREQuency:OFFSet
<Offset>
This command defines a frequency offset.
If this value is not 0
Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also
Parameters:
<Offset>
Range:
-1 THz to 1 THz
*RST:
0 Hz
Default unit: HZ
Example:
FREQ:OFFS 1GHZ
Manual operation:
See
[SENSe:]FREQuency:SPAN
<Span>
This command defines the frequency span.
If you set a span of 0
Hz in the Spectrum application, the R&S
FSV/A starts a mea-
surement in the time domain.
Setting basic measurement parameters