
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
932
User Manual 1178.8520.02 ─ 08
In case of high speed measurements, the resolution bandwidth has to be identical for
all ranges.
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the measurement range.
Parameters:
<RBW>
Resolution bandwidth.
Refer to the data sheet for available resolution bandwidths.
*RST:
30.0 kHz
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:VIDeo
<VBW>
This command defines the video bandwidth for a SEM range.
In case of high speed measurements, the video bandwidth has to be identical for all
ranges.
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the measurement range.
Parameters:
<VBW>
Video bandwidth.
Refer to the data sheet for available video bandwidths.
*RST:
10.0 MHz
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list.
Suffix:
<sb>
.
1 to 8
<ri>
irrelevant
Return values:
<Ranges>
Number of ranges in the sweep list.
Usage:
Query only
Configuring and performing measurements