Status Indicator RAPs
016-362 Cont UI Fail-2 RAP
BSD-ON:
BSD 16.1 - ESS
Cont UI Diagnostic Fail-2. UI Communication error has occurred. An error occurred during W/
R/V test of VRAM. The Command/Status line is normal if entering the Long Boot Mode.
Procedure
Turn the power OFF then ON. If the problem persists, perform the following:
1.
Disconnect and connect the UI cable (
PL 18.2
).
2.
Remove and re-install the UI I/F PWB (
PL 1.7
).
3.
Check the sw version of the controller sw - update if required
4.
Replace the UI cable (
PL 18.2
).
5.
Replace the UI I/F PWB (
PL 1.7
).
6.
Replace the Control Panel Assembly (
PL 18.3
).
7.
Replace the ESS PWB (
PL 35.2
).
016-364 Cont USB2.0 Host Fail RAP
BSD-ON:
BSD 16.1 - ESS
Cont USB2.0 Host Diagnostic Fail. An error occurred during W/R/V test of PCI Config interval
or W/R/V test of internal register.
Procedure
Turn the power OFF then ON. If the problem persists, replace the ESS PWB (
PL 35.2
). Check
the sw version of the controller sw - update if required.