5. Check that the arterial and venous blood lines has been
completely disconnected from the patient. Con
fi
rm this by
pressing the
Select key
for 3 seconds and the menu will
change to the
DISMOUNT OLD BLOOD CIRCUIT
menu.
When patient disconnection has been con
fi
rmed, the air detector
will be deactivated and the dialysis
fl
uid will automatically be
bypassed from the dialyzer.
6. Disconnect the dialysis
fl
uid tubes from the dialyzer and
connect them to the safety couplings of the machine. Remove
the dialyzer and the blood lines; follow the instructions
in "Machine aftercare", page 4:63 in part 1, point 25 and
additional warning.
7. Attach new dialyzer and blood lines; follow the
corresponding instructions in "Attach Disposables", starting
8. Press the
Select key
when new disposables have been
attached, the
Priming button
lights up indicating that priming
mode has been activated.
9. Prime and de-air the new blood circuit as usual procedure.
Follow the corresponding instructions starting on page 4:40
for "Manual Priming and Rinsing Procedure" and on page
4:46 for "Assisted Priming and Rinsing Procedure" in part 1.
Resume Treatment
10. Check/adjust the treatment parameters for the continuing treatment
(see "Always Check/Adjust", page 4:52) in part 1.
11. Connect the blood lines and start the treatment as usual procedure.
See corresponding instructions in "Connect the Patient", page 4:53
and "Start the Treatment", page 4:56 in part 1.
Especially note that
all treatment parameters
, set values as well
as accumulated values, from treatment start up until where it was
interrupted
have been saved
by the machine. If the Diascan or
Pro
fi
ling functions have been previously activated, attention alarms
will be generated to give the operator instructions on how to set/adjust
the values for the continuing treatment.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Operating the Machine
3:35
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...