Acetate concentrate
1.
Attach Acetate concentrate
1.a.
Remove the blue concentrate connector from the
stand-by port of the machine (compress the connector
before pulling the connector out).
1.b.
Attach the connector to a concentrate pick-up tube;
insert the connector into the pick-up tube port and push
it into place until it clicks in. Make sure that it clicks
in properly to prevent air leakage during treatment.
1.c.
Put the pick-up tube into the proper acetate concentrate
container (the red concentrate connector must remain
in the stand-by port).
2.
Con
fi
rm concentrates
The con
fi
rm concentrate menus will automatically appear
when the blue concentrate connector is removed from the
stand-by port.
2.a.
Select
ACETATE
using the
Keypad
2.b.
Con
fi
rm that the connected concentrates correspond
with the displayed concentrate alternative on the
Information Display, using the
Select key
.
Note that at any time after con
fi
rming the concentrates, it
is possible to enter the concentrate menus to check/select
concentrates. Press the
Fluid path button
and select
CONC
. Go
into the sub menus and check/select the connected concentrates
using the
Keypad
.
4:12
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...