PRIMING VOLUME LIMIT ACHIEVED
Check blood line. Press SELECT
Appears:
When the extra priming volume limit
has been reached.
This attention alarm can be removed by preset.
Machine actions:
None.
Measures:
If this attention alarm appears during initiation of the
treatment, check that the venous blood line is properly
inserted into the priming detector.
Press
Select key
to con
fi
rm.
Mute time:
2 min
Buzzer sound:
Soft buzzer
14:104
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order