Power supply
Mains voltage
115 V (at 50 Hz or 60 Hz) or 230 V (at
50 Hz or 60 Hz) (
∼
AC), (±12 %).
The voltage has to be speci
fi
ed before
installation.
Note!
The user must verify the quality of the
protective earth in the installation.
Protection class
Machine: Class 1 type B
BPM: Type BF
Power consumption
Max. 2025 W at 230 V
Max. 1575 W at 115 V
Cable
3 conductor cable, Length max. 3.5 m
Ratings:
230 - 250 V AC 13 -16A or
110 - 125 V AC 15 A
External fuses
For 115/230 V, 2 x 12AT for heater.
Mains plug
Earthed plug, 250 V AC / 10 -16 A,
approved or Hospital grade, earthed plug,
125 V AC / 15 A, approved.
Earth leakage current
max 500
μ
A
Patient leakage current
max 100
μ
A AC
max 10
μ
A DC
All leakage currents speci
fi
ed are without external equipment
connected to the AK 96.
9:12
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx