Note
•
Changes in treatment parameters (above all blood
flow rate) affect measured clearance (K) values
obtained during treatment and consequently, the
calculated Kt/V result. If the Kt/V calculation at
treatment end is based on few clearance results, or
only one, check that these results are relevant for the
complete treatment. This is in order to ensure that the
Kt/V value is a relevant one.
Note
12:30
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx