Rinse/Drain Program - Automatic Start
The machine is equipped with an automatic start function for the
rinse/drain program. This is a feature that enables the machine to
automatically start a rinse/drain program at a predetermined time. The
operator decides (programmes) which day of the week and what time
of the day the machine shall be
fi
nished, after having performed a
rinse/drain program.
The machine can be programmed and the function activated, during
any phase when the machine is switched on. Whenever this function
is activated, and the machine has been switched off, the rinse/drain
program will be started at the programmed time. See below for
programming and activation instructions.
Note
•
The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
Check whenever activated
That the dialysis
fl
uid tubes are connected to the safety couplings
and that the latches of the BiCart holder are closed.
That the points in the list "Check Before Switching the Machine
On" in chapter 4, part 1 on page 4:2 are ful
fi
lled.
That the concentrate connectors are placed in the corresponding
stand-by ports of the machine (blue and red), and that the yellow
disinfectant connector is placed in the parking port (marked P).
Programming and Activation Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
1. Press the
Rinse/Disinfection button
brie
fl
y.
2. The
RINSE/DISINFECTION MENU
will be displayed on
the Information Display.
3. Select
RINSE
.
8:36
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...