Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display. Then press
the
Select key
to open the venous pressure menu.
It is also possible to directly view the venous pressure
menu by pressing the lit
Venous Pressure button
.
The alarm limit setting of which the alarm applies
for will be open.
-
Check cause (the patient's blood access, the
extracorporeal circuit) and do necessary actions in
accordance with this. If the blood pump intermittently
starts and stops, consider adjusting the blood
fl
ow
rate or if necessary, manually stop the blood pump,
using the
Blood Pump buttons
.
-
If possible, adjust the venous pressure high alarm
limit using the
Key pad
.
After the venous pressure alarm has been attended to, and
if the venous pressure alarm window
is set to be wider
than the preset value, the
Venous Pressure button
will
start to
fl
ash. Press the
fl
ashing button to centralize the
alarm limits around the current value.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms
13:15
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...