BLOOD DETECTED IN ASSISTED PRIM
Appears:
When the priming detector detects blood during assisted
priming procedure.
Machine actions:
The blood pump stops.
Measures:
First, make sure that the patient has not been connected to
the blood lines. If this is not the case, clean the priming
detector lens. Alternatively call the authorized technician.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms