Continue to follow the instructions in chapter 4, "Hemodialysis -
Double Needle Treatment" starting on page 4:1 up to "Connect the
Patient", page 4:53 in part 1.
Connect the Patient
Note
•
The
CONCENTRATE STANDBY MODE
menu, displayed on
the Information Display, indicates that concentrate stand-by
mode is active. Resume dialysis
fl
uid preparation (deactivate
concentrate stand-by mode) by pressing the
Select key
for three
seconds, as the menu requests. The dialysis
fl
uid will be ready
within approximately 2 minutes.
•
Always follow the manufacturer's recommendations for the
minimum rinsing volume (for the blood lines and the dialyzer).
Note
After
CONNECT PATIENT
has been selected a connect patient
volume
will be measured by the machine once the blood pump has
been started. If blood has not been detected in the venous blood line
when this connect patient volume has been achieved, an alarm will
be generated;
Blood is not detected
. See “Connect Patient
Alarm”Connect Patient Alarm, page 13:49 in part 3 for instructions.
1. Check that the venous expansion chamber is in an upright
position and is
fi
lled to a correct level (refer to the
instructions for use of the individual blood line).
2. Check that there is no air in the blood lines and that the level
in the venous drip chamber is well above the air detector
head.
3. Manually activate the air detector alarm function by pressing
the
fl
ashing
Air detector button
if it has not previously been
activated during priming. If the air detector is not activated,
it will automatically be activated when the priming detector
detects blood. The activated air detector alarm function is
indicated by an unlit button.
4. Check that the blood pump has been stopped, if not, press
the
Blood pump button
. As
CONNECT PATIENT
has
been previously selected, the blood pump will start at
100
ml/min when the
fl
ashing
Blood pump button
later
is pressed. If desired, adjust the blood
fl
ow rate using the
Blood Pump Up/Down keys
to the proper value for treatment
start.
5. Press the
Blood Path button
.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Single Needle Treatment
5:5
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...