The change of UF rate for the graph in intervals is determined by
the following settings; total UF volume, treatment time, number of
intervals in the graph and start UF rate. Based on these parameters, the
machine will perform adjustments of the UF rate at intervals during the
treatment time when pro
fi
ling has been activated. See the following
fi
gure.
Figure 7:4
Graph in intervals for UF rate pro
fi
ling
For all the UF rate graphs, the operator is requested to readjust the
TMP alarm limits during treatment in accordance with the UF rate
and the dialyzer UF coef
fi
cient.
7:4
AK 96
®
Operator's Manual - Profiling
HCEN12239 Revision 12.2010
Program version 3.xx