TMP LIMITS TOO WIDE
Press SELECT key to confirm
Appears:
The TMP alarm window
is wider than the preset value.
Machine actions:
None.
Measures:
Press the
Select key
to centralize the alarm limits around
the current value. This text will appear at any time
during treatment when the blood pump has been started
or stopped, the ultra
fi
ltration has been activated or
deactivated, or when changes of treatment time and UF
volume have been made.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:115
Attention alarms in alphabetical order