Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display. Then press
the
Select key
to open the TMP menu.
It is also possible to press the lit
Fluid Path button
,
select
TMP
and go into the TMP menu, using the
Keypad
.
The alarm limit setting of which the alarm applies
for will be open.
-
Check cause (the patient for unaccountable weight
loss, the patient's blood access, the extracorporeal
circuit, the dialyzer) and do necessary actions in
accordance with this.
-
Do standard patient checks for incorrect weight loss
such as measuring blood pressure. If necessary,
discontinue treatment and call an authorized
technician.
-
If possible, decrease the set treatment time or increase
the set UF volume.
-
If possible, adjust the low TMP alarm limit using the
Key pad
.
When the Transmembrane pressure con
fi
rm menu
is displayed on the Information Display, press the
Select key
to con
fi
rm the settings i.e. that the current
TMP value and the automatically set TMP alarm limits
are set in accordance with the current UF rate and the UF
coef
fi
cient of the dialyzer being used.
13:40
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...