During the period of time when the measurement check
is being performed,
DIASCAN
, shown at the activity
fi
eld
of the Treatment Overview Menu, will be
fl
ashing.
When the measurement check has been performed, an
attention alarm text without a buzzer will appear on the
Information Display . Press the
Select key
to enter
the
LOG
menu. The measurement result values are also
displayed in the Diascan viewing menu.
When the measurement check has been performed
Diascan will automatically be turned to
OFF
. It is possible
to repeat the procedure at any time during treatment.
Note
•
Changes in treatment parameters (above all blood
flow rate) affect measured clearance (K) values
obtained during treatment and consequently, the
calculated Kt/V result. If the Kt/V calculation at
treatment end is based on few clearance results, or
only one, check that these results are relevant for the
complete treatment. This is in order to ensure that the
Kt/V value is a relevant one.
•
To ensure a correct result a single clearance
measurement check should be activated no later than
30 minutes before discontinuing treatment. This is to
have enough time to obtain a measurement result.
Note
12:26
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...