LOW FLOW HEAT ABORTED
Press SELECT key to confirm
Appears:
When the WRO 300 H cannot deliver hot water during
integrated heat disinfection program with WRO 300 H.
The low
fl
ow heat phase (where the machine receives hot
water from the WRO 300 H) will not start.
It can also appear during manual start procedure
of the low
fl
ow heat phase when the attention
alarm
Start low flow heat in WRO
,
Press SELECT key to continue
, has not been
con
fi
rmed within 5 minutes
.
Machine actions:
The low
fl
ow heat phase has automatically been
cancelled.
Measures:
Press the
Select key
to con
fi
rm the attention alarm. The
machine now enters drain phase as the circulation phase
(for information, see in "Heat Disinfection - General" on
page 8:9 in part 1) of the heat disinfection program is
complete. Call for an authorized technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Supersoft buzzer
14:90
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order