
6. Set the required blood
fl
ow rate suitable for the patient using
the
Blood Pump Up/Down keys
; the blood
fl
ow menu will be
displayed, adjust
SET QB
.
When the priming detector detects blood in the venous blood
line, the blood path of the Flow Diagram lights up red and
the treatment time starts to count down on the Time Display.
When blood has been detected for 20 seconds the blood path of
the Flow Diagram will remain lit until treatment end has been
con
fi
rmed (by attention alarm) and the machine no longer detects
blood in the venous blood line. This is valid even if the venous
blood line is removed from the priming detector during treatment
(i.e. the blood alarms will be kept).
The Treatment Overview Menu will start to be displayed and
continues to be displayed at all times during treatment when
blood is detected. See in "Overview Displays", page 3:23 in
part 1 for details.
The
Venous Pressure
and
Arterial Pressure button
will be
fl
ashing.
The
UF Start/Stop button
will be
fl
ashing.
See "Start the Treatment" in the next following section for
information on how to continue.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:55
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...