FLUID TUBE SENSOR TEST
Remove tubes from safety bypass
Appears:
When a disinfection or rinse program has been activated
and the dialysis
fl
uid tube sensor test has been interrupted.
Machine actions:
None.
Measures:
Remove and attach the dialysis
fl
uid tubes from the safety
couplings of the machine to restart the test.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:59
Attention alarms in alphabetical order