Make sure the blood pump segment collars are outside the
pump housing.
If manually threaded
Place and hold the blood pump segment in the way shown
in the corresponding
fi
gure.
Press on the blood pump segment at the point shown
in the
fi
gure before turning the blood pump rotor in an
anti-clockwise direction. This is to hold the segment collar
outside the pump housing while threading it.
Continue turning the blood pump rotor in an anti-clockwise
direction, while holding the segment, to thread the segment
through the blood pump. The pins on the blood pump rotor,
pointed out as
in the
fi
gure, hold the segment in place.
4:20
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx