6. Select
VOLUME
using the
Keypad
.
The distribution volume is now
to be set to a correct value for
the individual patient. If desired,
it is possible for the machine to
carry out a distribution volume
calculation using the Watson
formula, see "Distribution
If the value is to be set manually,
when
VOLUME
has been
selected, set a proper value using
the
Display Up key
. 40 l is the
default starting point value .
Adjust the value as necessary.
Note that the distribution volume
has to be properly estimated
for the individual patient. The
calculation of Kt/V done in the
machine is entirely dependent
on the value for the individual
patient being the correct
one. For further information;
see "Glossary of Diascan®
parameters and key terms", page
7. To set the alarm limit
for minimum Kt/V, select
Kt/V LIMIT
and set a proper
alarm limit value
using the
Keypad.
.
12:28
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010