HIGH VENOUS PRESS TEST NOT MADE
Apply ven press >50mmHg for test
Appears:
During priming or when treatment has been started and
the machine has not been able to test the venous pressure
measurement function.
Machine actions:
10 minutes after blood has been detected the treatment
will stop if the attention alarm is left unattended.
Measures:
Establish a stable venous pressure over 50 mmHg until
the attention alarm disappears.
Mute time:
2 min.
Buzzer sound:
Continuous buzzer 5 min after blood detection
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:75
Attention alarms in alphabetical order