3. Select
HEAT
.
4. Select
SET AUTO HEAT
.
5.
AUTO HEAT MENU
will be displayed.
6. Check that
TIME NOW
is correct. Change if required.
7. Set
READY DAY
i.e. the day when the program is to be
fi
nished.
8. Set
READY TIME
i.e. the time when the program is to be
fi
nished.
9. If the heat disinfection program should be performed other
weekdays repeat point 7 to 9 for every day. The auto heat
function can be switched off for the selected day by setting
the ready time to
OFF
.
10. Press
Select key
3 seconds to activate the auto heat function
when all days are set.
11. The machine can now be switched off using the
On/Off button
. It will automatically start and initiate the set
program to have it completed in due time.
The
Rinse/Disinfection button
will always be lit when the auto
heat function has been activated and the machine is switched on.
If changes are to be made or the auto heat process shall be
inactivated, enter the
AUTO HEAT MENU
as described above.
The
Select key
is used to activate and deactivate the auto heat
function.
Should days in the schedule be deactivated the set “hours” in
READY TIME
for the intended day is set to
OFF
.
In combination with CleanCart
®
cartridge
If a heat disinfection program with CleanCart cartridge is desired,
place a CleanCart cartridge into the BiCart holder before the automatic
start of the heat disinfection program is due to start. Remove the empty
CleanCart cartridge and close the latches of the BiCart holder when the
program is
fi
nished before restarting the machine.
8:24
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...