BATTERY FAILURE
Change battery or press SELECT
Appears:
When the battery that supplies the power failure buzzer
is not properly charged. It is only displayed when blood
is not detected.
Machine actions:
None.
Measures:
-
The battery needs to be changed or recharged.
Contact an authorized technician.
-
Press the
Select key
to con
fi
rm the attention alarm
and to be able to start the treatment.
If the attention alarm is con
fi
rmed without changing or
recharging the battery, the machine will not be able to
alarm or restart in case of power failure.
Mute time:
Permanently muted buzzer
Buzzer sound:
Soft buzzer (during function check)
No buzzer (during disinfectant program)
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:11
Attention alarms in alphabetical order