General
Application
The Diascan function can be used for:
•
Clearance measurement (K)
in order to read off the
current dialysis ef
fi
ciency for the ongoing treatment.
•
Dialysis dose measurement (Kt or Kt/V)
in order
to check that the prescribed dialysis dose is being
maintained, which provides quality assurance for the
treatments.
•
Kt/V target supervision
, which means that the
machine constantly calculates if the desired and
minimum values for Kt/V, set by the operator for
each individual patient, can be reached at treatment
end. If not, the operator will be immediately noti
fi
ed
during treatment.
The measurement checks can be performed one at a time
when required, or continuously at set intervals during
treatment.
If Kt/V measurement is required, distribution volume
(
VOLUME
) has to be set before the measurement check
takes place. The distribution volume is patient-related
(based on the patient's dry weight) and has to be properly
estimated and set by the machine operator in order to
obtain a correct Kt/V value.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Diascan
12:3
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...