FLUID PATH OBSTRUCTION
Check fluid tubes and dialyzer
Appears:
The machine has detected an obstruction in the
fl
uid path.
Machine actions:
None.
Measures:
Check if the obstruction is external i.e. if it involves
the dialysis
fl
uid tubes or the dialyzer. This is done by
pressing the
Fluid bypass button
. If the attention alarm
disappears, the obstruction is external. Check dialysis
fl
uid tubes and dialyzer for possible obstruction.
If the attention alarm does not disappear, the obstruction
is internal. Call an authorized technician.
Mute time:
2 min.
Buzzer sound:
Continuous buzzer
14:56
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order