HIGH INLET WATER COND.
#.#mS/cm
Confirm low K+, press SELECT key
Appears:
Appears during function check if the inlet water
conductivity is between 0.5 and 2.5 mS/cm. The
appearance of the attention alarm can be preset.
Machine actions:
Depending on the inlet water conductivity, minor
electrolytes concentration (potassium, calcium,
magnesium) will be reduced by 4 - 20% in the dialysis
fl
uid.
Measures:
When this attention alarm appears, the operator must
consult the attending physician before the attention alarm
is con
fi
rmed and a treatment is started. It will not be
possible to start a treatment until the attention alarm is
con
fi
rmed.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:73
Attention alarms in alphabetical order