SET DIA FLUID FLOW NOT REACHED
Readjust flow or call service
Appears:
Current dialysis
fl
uid
fl
ow rate
(
ACTUAL DIA FLUID FLOW
) does
not correspond with the set value
(
SET TREATMENT DIA FLUID FLOW
). It is not
possible for the machine to set the required dialysis
fl
uid
fl
ow rate.
Machine actions:
None.
Measures:
If it is possible for the treatment to continue with a lower
fl
ow, adjust the set value to correspond with the current
value; the attention alarm will disappear.
If this is not possible, consult an authorized technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:111
Attention alarms in alphabetical order