Blood Leak Alarm
Appears:
Blood has entered the
fl
uid path of the machine
downstream the dialyzer.
Machine actions:
-
The blood pump stops.
-
The arterial blood line clamp closes.
-
The ultra
fi
ltration rate is set to zero and the dialysis
fl
uid is bypassed from the dialyzer (when blood is
detected).
-
The machine indicates the alarm by a
fl
ashing part of
the
fl
uid
fl
ow path.
Possible cause:
-
A dialyzer leakage, resulting in blood
fl
owing
from the blood compartment to the dialysis
fl
uid
compartment, has occurred.
13:34
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx