Alarms - General
Alarm Indication
When an alarm is generated, the machine indicates the
alarm as follows:
-
the
Alarm button
is
fl
ashing
-
the
Mute button
is
fl
ashing
-
the buzzer sounds continuously
-
the button (not valid for Blood
Leak alarm and Technical
Alarm) of which the treatment
parameter is connected to, and
the alarm applies for, lights up.
See the description for each
alarm in the "Alarm List" page
General Alarm Handling
Press the
fl
ashing
Alarm button
. The button lights
up and the alarm text (which gives information on
what has triggered the alarm) is displayed in the alarm
list of the Information Display. At the same time as
the
Alarm button
is being pressed, the buzzer will be
simultaneously muted for 2 min.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms
13:5