CHEMICAL FILLUP NOT SUFFICIENT
Check supply of disinfectant
Appears:
During chemical disinfection program, in
fi
ll-up phase.
Machine actions:
The disinfection program is interrupted.
Measures:
Check the pick-up tube and the disinfectant container.
Machine retries the
fi
ll-up phase.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Supersoft buzzer
14:28
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order