DECALCIFICATION REQUIRED
hours since last decalcific.###
Appears:
When a decalci
fi
cation program is needed.
Machine actions:
None.
Measures:
If possible, run a decalci
fi
cation program, i.e. a heat
disinfection program with CleanCart-C cartridge or liquid
citric acid. If not, the attention alarm can be cleared by
pressing the
Select key
. If the machine is not disinfected
after treatment, the attention alarm will reappear during
the next treatment.
The period of time between the last performed
decalci
fi
cation and the appearance of the attention alarm
can be preset.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Supersoft buzzer
14:34
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order