Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display.
-
Check cause (the patient's blood access, the
extracorporeal circuit, infusions) and do necessary
actions in accordance with this.
-
Adjust the level in the venous drip chamber by using
the level adjustment knob. Turn the level adjustment
knob anticlockwise to raise the level and clockwise
to lower it.
-
If necessary, override (bypass) the machine alarm
actions by pressing the lit
Air Detector button
for 3
seconds. The blood pump will start up at 100 ml/min
blood
fl
ow rate, adjust if necessary. If the cause of
the alarm has not been corrected within the override
time , the machine alarm actions will automatically
recur. If necessary, repeat the override procedure.
Especially note that the blood
fl
ow rate will remain
at 100 ml/min after the override period and has to
be manually adjusted. See also "General Alarm
Handling", page 13:5.
13:20
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx