IMPOSSIBLE AUTO START SETTING
Check auto heat/rinse ready time
Appears:
When the machine is switched off and the time for when
the heat disinfection program or rinse program is set to
be
fi
nished is too close to present time.
Machine actions:
The machine will not be switched off until the time
setting has been changed.
Measures:
Change the time setting for when the rinse program or
heat disinfection program is set to be
fi
nished.
Mute time:
2 min.
Buzzer sound:
Soft buzzer