Heat Disinfection Program - Automatic Start
Note
•
The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
The machine is equipped with an automatic start function for the heat
disinfection program. This is a feature that enables the machine to
automatically start a heat disinfection program at a predetermined time.
The operator decides (programmes) which day of the week and at what
time of the day the machine will be
fi
nished after having performed a
heat disinfection program.
The machine can be programmed and the function activated, during
any phase when the machine is switched on. Whenever this function is
activated, and the machine has been switched off, the heat disinfection
program will be started at the programmed time. See below for
programming and activation instructions.
The machine must be connected to the water supply as well as to
power, when performing a heat disinfection program with automatic
start. The drain tube must be connected to the drain.
Check whenever activated
That the dialysis
fl
uid tubes are connected to the safety couplings
and that the latches of the BiCart holder are closed.
That the points in the list "Check Before Switching the Machine
On" in chapter 4, part 1 on page 4:2 are ful
fi
lled.
That the concentrate connectors are placed in the corresponding
stand-by ports of the machine (blue and red), and that the yellow
disinfectant connector is placed in the parking port (marked P).
Programming and Activation Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
1. Press the
Rinse/Disinfection button
brie
fl
y.
2. The
RINSE/DISINFECTION MENU
will be displayed on
the Information Display.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Hygiene and Maintenance
8:23
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...