If the measuring process needs to be
interrupted, press the
Select key
to
stop the procedure.
If desired, go back to the former
menu using the
Back key
.
BPM
will
be
fl
ashing at the activity
fi
eld of the
Treatment Overview Menu during
the ongoing measurement check.
If desired, it is possible to
preset an attention alarm that
will advise the operator to
check the result values in the
*** BPM LOG LIST ***
every
time a measurement has been
performed. This attention is optional
and must be preset by a service
technician. The list starts with the
latest measurement at the top. See
"BPM Log List", page 11:6.
The latest measurement result
values will also be displayed on the
BPM MENU
.
TIME
shows current time for when
the blood pressure measurement
check was completed.
SYS
/
DIA
shows systolic/diastolic
blood pressure.
MEAN
shows mean blood pressure.
The value beside
PULSE
shows
pulse rate.
To read previous measuring results; select
LOG
using the
Keypad
to open the
*** BPM LOG LIST ***
. See
"BPM Log List", page 11:6.
11:8
AK 96
®
Operator's Manual - BPM
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...