Appears:
The arterial pressure has passed the set arterial pressure
low alarm limit.
Machine actions:
-
The blood pump stops.
-
The venous blood line clamp closes.
-
The ultra
fi
ltration rate is set to zero.
When the arterial pressure is within the set alarm limits
again, the blood pump automatically starts and the venous
blood line clamp opens.
Possible cause:
-
The performance of the patient's vascular blood
access is not in accordance with the set blood
fl
ow
rate.
-
The arterial blood line is kinked or clotted between the
arterial needle and the arterial pressure measurement
point.
-
The position of the arterial needle has changed.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms