SAGS TEST FAILED
Check water supply
Appears:
During disinfection program, when the machine cannot
maintain the pressure required at the dialysis
fl
uid tube
connectors.
Machine actions:
The disinfection program stops.
Measures:
Check if the AK 96 dialysis machine is properly
connected to the water supply.
Mute time:
Permanently muted buzzer.
Buzzer sound:
No buzzer
14:110
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order