Air bubbles or particles in the
fl
uid path of the
machine, which can cause a false blood leak alarm,
have entered the blood leak detector.
Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display.
-
Check the machine inlet dialysis
fl
uid tube (the
tube to the machine from the dialyzer) for blood.
Follow the routines of the clinic which may differ
depending on the amount of blood has leaked to the
fl
uid compartment and the status of the patient. The
measures may vary between immediate discontinuing
of the treatment without returning the blood to the
patient, to continuing treatment without action.
-
The blood leak machine alarm actions may be
overridden (bypassed) by pressing the
Select key
.
The blood pump will start up at 50 ml/min blood
fl
ow rate, adjust if necessary. The blood leak detector
will automatically be rinsed with dialysis
fl
uid during
the override period. If the cause of the alarm has not
been corrected within the override time, the machine
alarm actions will automatically recur. If necessary,
repeat the override procedure. Especially note that
the blood
fl
ow rate will remain at 50 ml/min after the
override period and has to be manually adjusted. See
also "General Alarm Handling", page 13:5.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms
13:35
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...