Schedule for Hygiene and Maintenance
The schedule below is recommended for use by the operator/user of the
AK 96 dialysis machine. This recommendation is to keep a high level
of performance of the machine. It is also a guideline for maintaining
the hygiene of the
fl
uid path and the exterior of the machine to ensure
the safety of the patient.
Frequency
Measures
Cleaning
Decalci
fi
-
cation
Disinfec-
tion
After each
treatment
1) Heat disinfection program.
2) Wipe down the outside of the machine
with 70% ethanol or 60% isopropanol.
x
At least after every
third treatment
Combine the heat disinfection program
with CleanCart-C cartridge or liquid
citric acid.
x
x
x
At least once every
seventh treatment
day
Heat disinfection program in
combination with CleanCart-A
cartridge or a chemical disinfection
program with sodium hypochlorite.
This should be performed directly
after a heat disinfection program in
combination with CleanCart-C cartridge
or liquid citric acid.
x
x
x
When time after
disinfection has
exceeded seven
days
Heat disinfection program before
treatment.
x
Once every one
to three months if
UFD is installed
1
Change of ultra
fi
lter followed by a heat
disinfection program.
x
1
For more information, see "Ultra
lter - Frequency of Change" in chapter 1 part 1.
8:6
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...