Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display. Then press
the
Select key
to open the arterial pressure menu.
It is also possible to directly view the arterial pressure
menu by pressing the lit
Arterial Pressure button
.
The alarm limit setting of which the alarm applies
for will be open.
-
Check cause (the patient's blood access, the
extracorporeal circuit) and do necessary actions in
accordance with this. If the blood pump intermittently
starts and stops, consider adjusting the blood
fl
ow
rate or if necessary, manually stop the blood pump,
using the
Blood Pump buttons
.
-
If possible, adjust the arterial pressure low alarm limit
using the
Key pad
.
-
If necessary, override (bypass) the machine alarm
actions by pressing the lit
Arterial Pressure button
for
3 seconds. The blood pump will start up at the set
blood
fl
ow rate, adjust if necessary. If the cause of
the alarm has not been corrected within the override
time, the machine alarm actions will automatically
recur. If necessary, repeat the override procedure.
See also "General Alarm Handling", page 13:5.
13:12
AK 96
®
Operator's Manual - Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...