Acidic concentrate and BiCart
®
cartridge
1.
Attach Acidic concentrate
1.a.
Remove the red concentrate connector from the
stand-by port of the machine (compress the connector
before pulling the connector out).
1.b.
Attach the connector to a concentrate pick-up tube;
insert the connector into the pick-up tube port and push
it into place until it clicks in. Make sure that it clicks
in properly to prevent air leakage during treatment.
1.c.
Put the pick-up tube into the proper acidic concentrate
container (the blue concentrate connector must remain
in the stand-by port).
If SoftPac container for acidic concentrate is being used, connect
the red concentrate connector to the particular connector used for
SoftPac containers and insert it into the SoftPac container.
2.
Attach BiCart cartridge
2.a.
Open the upper latch of the BiCart holder, pull out the
latch
and lift upwards to its uppermost position
.
Open the lower latch
.
2.b.
Place the bottom port of the BiCart cartridge into the
hole of the lower latch as shown in the corresponding
fi
gure.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:9
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...