Wait 30 seconds to bypass
dialysis fluid
Appears:
If the
Fluid Bypass button
is pressed during the
fl
ushing
of the dialysis
fl
uid tubes.
This attention alarm can be removed by preset.
Machine actions:
Automatic
fl
ush of the
fl
ow path where the ultra
fi
lter is
placed is performed.
Measures:
Wait until the attention alarm has disappeared and the
bypass path of the Flow Diagram lights up green before
bypassing the dialysis
fl
uid.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
14:132
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order