The AK 96 dialysis machine must be turned off before the blood leak
detector is cleaned.
Unscrew the blood leak detector cover and clean the lenses with a soft
non-
fl
uffy cloth, moistened with disinfectant solution.
Note
•
Make sure that the sealing ring on the inside of the cover is
securely in place when replacing the cover.
Note
BPM Cuff
Cleaning instructions for the BPM cuff can be read in the "package
insert" lea
fl
et which is enclosed to the cuff upon delivery.
8:42
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx