Attach tubes to safety bypass
Press SELECT to start disinf
Appears:
When it is not possible for the machine to detect that the
dialysis
fl
uid tubes are attached to the safety couplings.
The corresponding machine test can not be made.
Machine actions:
The disinfection or rinse program has been interrupted.
Measures:
To be able to continue the activated disinfection or rinse
program; check that the dialysis
fl
uid tubes are properly
attached to the safety couplings of the machine. Con
fi
rm
this by pressing the
Select key
and the activated program
will be resumed.
Alternately, consult an authorized technician.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:10
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order