Heat disinfection program including WRO 300 H
UFD not installed
UFD installed (option)
Temperature
93 °C
93 °C
Fill up phase
10
13
Circulation phase
15
15
Low
fl
ow heat phase
20
20
Drain phase
4
4
Total time
49
52
Note
•
Drain phase follows after low heat phase. WRO 300 H will
start heat disinfection simultaneously (if preset, see WRO 300
H Service Manual).
Note
Disinfection and Cleaning - Rinse/Drain
Note
•
Total time for disinfection programs is estimated and may vary.
Note
Rinse/Drain
10 min.
Drain
4 min.
Disinfection and Cleaning - Exterior Cleaning
All outside parts of the machine can be cleaned with ethanol (70 %) or
isopropanol (60 %).
9:10
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx