8.n.
If desired, set a heparin pump stop time
STOP LIMIT
.
The stop time is the time in minutes at which the
heparin pump should stop to run before treatment end.
When the heparin solution
fl
ow rate (
FLOW RATE
) and/or a bolus
dose of heparin solution (
BOLUS
) has been set above zero, the heparin
pump will automatically start to run at treatment start when blood is
detected in the venous blood line. If a heparin solution bolus volume
BOLUS
has been set, this will be administered
fi
rst. After this the
heparin solution will automatically be administered at the set
fl
ow rate.
Note that the sum of accumulated bolus volume during priming and
total accumulated volume since treatment start will be shown below
ACC HEP
. To stop the heparin pump from running during treatment;
set
FLOW RATE
to zero and bolus volume (
BOLUS
) to zero.
The heparin pump will run concurrent to the blood pump in double
needle treatment. In single needle treatment it will be running
continuously irrespective of if the blood pump is running or not.
During machine set-up and priming the heparin pump can be started
at any time by pressing the
Select key
for three seconds until the text
Press SELECT 3 seconds to start bolus
disappears;
the heparin pump immediately starts to run. (However, there is one
exception, it is not possible to run the heparin pump when performing
recirculation priming). Note that to stop the heparin pump from
running during priming; set bolus volume (
BOLUS
) to zero.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:27
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...