DIALYSIS FLUID FLOW TOO LOW
Appears:
Appears during treatment when the dialysis
fl
uid
fl
ow
rate is too low. The dialysis
fl
uid
fl
ow rate has been set
below the low alarm limit
or there is an obstruction in
the dialysis
fl
uid
fl
ow path.
Machine actions:
None.
Measures:
If possible, increase the dialysis
fl
uid
fl
ow rate.
If this is not possible, consult an authorized technician.
It is possible to mute the buzzer by pressing the
Mute button
, but if the attention alarm is left without
taking measures, it will reappear the next treatment.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:35
Attention alarms in alphabetical order